negative-pressure, ventilated cabinet usu-
ally operated with an open front and a
mi
nimum face velocity at the work open-
ing of at least 75 ft/min. A Class I BSC is
si
milar in design to chemical fume hood
except all of the air from the cabinet is
exhausted through a HEPA filter (either
into the laboratory or to the outside).
Class II BSC: A ventilated
BSC that
protects personnel, product, and the
work environment. A Class II BSC has
an open front with inward airflow for per-
sonnel protection, downward HEPA-fil-
tered laminar airflow for product protec-
tion, and HEPA-filtered exhausted air for
environmental protection.
Type A1 (formerly, Type A): T
h
ese
Class II BSCs maintain a minimum in-
flow velocity of 75 ft/min, have HEPA-
fil
tered downflow air that is a portion
of the mixed downflow and inflow air
from a common plenum, may exhaust
HEPA-filtered air back into the labora-
tory or to the environment through an
e
x
haust canopy, and may have posi-
tive-pressure contaminated ducts and
pl
enums that are not surrounded by
negative-pressure plenums. They are
not suitable for use with volatile toxic
chemicals and volatile radionucleo-
tides.
Type A2 (formerly, Type B3): These
Class
II
BSCs maintain a minimum
inflow velocity of 100 ft/min, have
HEPA-filtered downflow air that is a
portion of the mixed downflow and
inflow air from a common exhaust
plenum, may exhaust HEPA-filtered
air back into the laboratory or to the
environment through an exhaust
canopy, and have all contaminated
ducts and plenums under negative-
pressure or
surrounded by negative-
pressure ducts and plenums. If these
cabinets are used for minute quanti-
ties of volatile toxic chemicals and
trace amounts
of radionucleotides,
they must be exhausted through
properly functioning exhaust cano-
pies.
Type B1: These Class II BSCs main-
t
ain a minimum inflow velocity of 100
ft
/min, have HEPA-filtered downflow
air composed largely of uncontami-
nated, recirculated inflow air, exhaust
mo
st of the contaminated downflow
air through a dedicated duct exhaust-
ed to the atmosphere after passing
it
through a HEPA filter, and have all
contaminated ducts and plenums un-
der negative pressure or surrounded
by
negative-pressure ducts and ple-
nums. If these cabinets are used
fo
r work involving minute quantities
of volatile toxic chemicals and trace
amounts of radionucleotides, the
work must be done in the directly ex-
hausted portion of the cabinet.
Type B2 (total exhaust): T
h
ese Class
II BSCs maintain a minimum inflow ve-
locity of 100 ft/min, have HEPA-filtered
do
wnflow air drawn from the laboratory
or the outside, exhaust all inflow and
downflow air to the atmosphere after
filtration through a HEPA filter with-
out recirculation inside the cabinet or
re
turn to the laboratory, and have all
contaminated ducts and plenums un-
der negative pressure or surrounded by
di
rectly exhausted negative-pressure
ducts and plenums. These cabinets
may be used with volatile toxic chemi-
cals and radionucleotides.
Hazardous Drugs in Health Care Settings 43