2
Polymer beads (Fluidigm, USA): Polymer beads laced with
lanthanide metals (
140
Ce,
151
Eu,
153
Eu,
165
Ho and
175
Lu) were
used as a suitable reference material due to the similarity in size
(2.5 µm) and density (1.05 g/cm
3
) to cells. The presence of
lanthanides in the beads enables them to be measured using
SC-ICP-MS. Although the size cut-off for the baffled cyclonic spray
chamber is in the 1-5 µm range,
2,3
the transport of these larger
beads (2.5 µm) is low, as illustrated later in Figure 2. The beads
were supplied at a concentration of 330,000 beads per mL and
diluted 10 fold for analysis. The mass of metal per bead can be
found in Table 2.
Gas and Sample Flows Optimization
The make-up and nebulizer gas flows and sample flow rate
had to be optimized to allow for equal transport of nano- and
micron-sized objects through the introduction system. Figure 1
shows the transport efficiency for nano- and micron-sized
objects through the introduction system for different sample
flow rates. It can be seen that as the sample flow rate decreases,
the transport efficiency for both nano- and micron-sized objects
increases and the difference between their transport gets less
pronounced with a transport efficiency of about 31% for both
the 60 nm Au NIST standards and the lanthanide-doped micron
beads (Fluidigm). The optimized conditions for this system can
be seen in Table 1.
Transport Efficiency Validation: Baffled Cyclonic vs.
Asperon Spray Chamber
A comparison study of the transport efficiency of nano- and
micron-sized particles between the baffled cyclonic and Asperon
spray chambers is shown in Figure 2. The systems were both
optimized for maximum intensities while keeping oxide and
double-charge formation below 2.5% (Table 1). It can be seen
in Figure 2 that both introduction systems transport a sufficient
amount of NPs (NIST 8013 60 nm Au NPs) into the ICP-MS to
provide a statistically significant measure of the mean and
standard deviation of the NPs (either in diameter or mass per NP)
and particle number concentration showing that both systems
work equally well for nanoparticles. However, the number of
micron-sized particles transported by the baffled cyclonic spray
chamber is low, with only a few beads being analyzed compared
to the Asperon spray chamber, where a significant number of
beads were counted.
Standards for SP-ICP-MS and SC-ICP-MS: Standards of 1, 2
and 3 ppb
140
Ce,
151
Eu,
153
Eu,
165
Ho and
175
Lu were prepared
for measuring the mass of metals in the polystyrene beads.
All standards for SC-ICP-MS were prepared in ultra-pure water.
The transport efficiency was determined with the 60 nm Au
NPs (NIST 8013) at a concentration of 50,000 part. mL
-1
.
Digestion of polystyrene beads: Digestion was accomplished
by placing 5 mL of the bead suspension into a PTFE digestion
vessel with 5 mL of hydrogen peroxide (Optima grade) and
10 mL of nitric acid (Optima grade). The mixture was allowed to
sit for 10 minutes to allow gasses to be released from any initial
reactions before the vessels were sealed. The beads were then
digested in a Titan MPS
™
Microwave (PerkinElmer), following
the program in Table 3. After they had cooled, the samples were
diluted to 2% acid for analysis. Standards of 10, 50, 100 and
200 ppb Lu, Eu, Ho and Ce were matrix-matched to the samples
for ICP-MS analysis. Ge and In were spiked into the samples as
internal standards.
Element Atoms Per Bead (± 15%) Mass (ag) (± 15%)
Ce 140 19.9E6 4626.26
Eu 151 11.3E6 2833.38
Eu 153 12.0E6 3048.75
Ho 165 7.6E6 2082.32
Lu 175 9.8E6 2847.82
Table 2. Supplier information on lanthanide laced polystyrene beads.
Stage
Temperature
(°C)
Pressure
(bar)
Ramp
(min)
Hold
(min)
P
(%)
1 150 70 2 5 60
2 190 75 2 10 80
3 210 80 2 15 90
4 50 80 1 10 0
5 50 0 0 0 0
Table 3. Conditions for microwave acid digestion of polystyrene beads.
Figure 1. Optimization of sample ow rates through the introduction system to maximize
the transport of nano- and micron-sized objects into the plasma.
0
10
20
30
40
0 500 1000 1500 2000 2500 3000
Transport efficiency %
Particle size (nm)
TE for nano- and micron-sized particles at
different sample flow rates
15.23
17.08
20.95
38.5
52.35