SECTIONING
10. Keep your samples in the cryostat until they are ready to be sectioned. Do not
allow the tissue to thaw at any time. It is best to only work with one sample at a
time.
11. During trimming, keep the antiroll device up. Advance the knife by turning the
handwheel and start trimming the block face. Continue trimming until the
specimen is exposed to the point desired for sections. Discard the first 2-3
sections to ensure that desired thickness is achieved.
12. Keep the knife and specimen free of debris by brushing upwards with a brush.
13. When ready to section, lower the antiroll device glass onto the stage in the
correct position. This will catch the sections as they are cut.
14. Continue to advance the knife to start taking sections. When a desired section is
cut, lift the antiroll plate and touch a glass slide face down onto the section. The
section should jump and stick onto the slide.
15. Multiple sections can fit on to a slide. Orientation and location of the tissue
should be considered before placing the tissue. If a section needs to be
repositioned in any way, a brush can be used to manipulate the section.
16. The slides can be stored in the cryostat while sectioning. Let the slides dry at
room temperature before staining.
SHUT DOWN
1. When you are done sectioning, lock the handwheel with the specimen holder in
the top position. Press the rewind button to move the specimen holder all the way
to the back, to the rear position.
2. Remove the blade from the knife holder and discard it in the sharps container.
3. Remove the chuck from the specimen holder and also remove the specimen
from the chuck. The specimen can also be removed with running warm water.
Clean the chuck with water at the sink or 70% ethanol and wipe the chuck with a
Kimwipe.
4. The cryostat chamber and accessories can be cleaned using 70% ethanol on a
Kimwipe.
5. Make sure the window is down and the chamber light is off. Leave the main
power on. Do not shut down the main power of the instrument.
STORAGE
Slides can be stored in -20°C if being used in the next few weeks and should be stored
in -80°C if they are going to be stored for months or years at a time. The slides should
only be thawed when they are ready to be used.
BLADES
Use with caution!! Microtome blades are very sharp and can cause serious injuries.
Never touch a blade with your fingers. Always lock the handwheel when manipulating
the blade. Do not clean the blade along its length. Wipe the blade from the back edge to
the cutting edge with a brush when cleaning. Put blade guard up when not sectioning.
Never leave the microtome unattended with an exposed blade in position. If you need to
step away, remove the blade or use the blade guard.