Welch Allyn ProBP 2400 digital blood pressure device
EN
1
ProBP 2400
Digital Blood Pressure Device
Directions for use
Advancing Frontline Care
powered by
CAUTION Federal law restricts this device to sale by or on the order of a physician or licensed healthcare provider
2
Introduction
This directions for use manual is a comprehensive guide designed to help you understand the capabilities and operation
of the ProBP 2400 digital blood pressure device. Read this manual thoroughly before attempting to setup, congure, use,
troubleshoot, or maintain the device.
Intended use
The ProBP 2400 is a non-invasive digital blood pressure device using oscillometric technique and an upper-arm blood
pressure cuff to measure systolic and diastolic blood pressures, pulse rate and mean arterial pressure (MAP) for use in
pediatric and adult populations with arm cuff circumference sizes ranging from 14 -52 cm.
The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading
once the irregular heartbeat is detected.
The device can accurately measure blood pressure in pregnant patients including those with known or suspected
preeclampsia.
The ProBP 2400 combines the advantages of an automatic blood pressure device and auscultatory sphygmomanometer. It
is designed to provide convenient, accurate and reliable ofce blood pressure measurements according to guidelines of the
European Society of Hypertension (ESH)
1
, American Heart Association (AHA)
2
, and World Health Organization (WHO)
3
with
the only modication that the ProBP2400 performs 3 repeated measurements always, regardless of the result of the rst two
measurements.
EN
3
Welch Allyn ProBP 2400 digital blood pressure device
1
Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure
measurement in humans: a statement for professionals from the subcommittee of professional and public education of the american heart association council
on high blood pressure research. Circulation 2005;111:697-716.
2
Whitworth JA. 2003 World Health Organization (WHO)/International Society of Hypertension (ISH) statement on management of hypertension. J Hypertens
2003;21:1983-92.
3
O’Brien E, Asmar R, Beilin L, et al. Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement.
J Hypertens 2005;23:697-701. E, Asmar R, Beilin L, Imai Y, et al. European Society of Hypertension recommendations for conventional, ambulatory
and home blood pressure measurement. European Society of Hypertension Working Group on Blood Pressure Monitoring. J Hypertens 2005;23:697-
701.2003,21:1983-1992.
Pregnancy
Approximately 20% of women develop hypertension during pregnancy (Pre-eclampsia or Toxemia) which may affect
pregnancy. Pre-eclampsia, can usually be recognized by a clear increase in blood pressure and high urine protein levels after
20 weeks of gestation. Since many oscillometric devices appeared to be unsuitable for use in pregnancy and Pre-eclampsia
health care authorities require that blood pressure monitors used for this vulnerable patient group are specically tested.
The Welch Allyn ProBP 2400 digital blood pressure monitor has successfully passed this validation and therefore may be
recommended for use during pregnancy and Pre-eclampsia.
4
EN
5
Welch Allyn ProBP 2400 digital blood pressure device
Symbols .................................................................................. 6
Warnings and cautions
General warnings and cautions .............................................. 7-8
NIBP (non-invasive blood pressure) warnings and cautions ...... 9
Product description
Name of parts and display .................................................. 10-11
Components and accessories ................................................. 12
Before using ProBP 2400
Attaching the power plug to the power adapter ........................ 13
Selecting Units of Measure ...................................................... 14
Selecting the correct cuff ......................................................... 15
Fitting the cuff properly ....................................................... 16-17
Taking measurements in Single Reading (1x), Manual and Three
Reading Average (
) Mode
«1x» Mode .............................................................................. 18
«Manual» Mode ................................................................. 19-20
«
» Mode .............................................................................. 21
Special Functions
MAP (Mean Arterial Pressure) ................................................. 22
Irregular heartbeat detector in «1x» Mode ............................... 22
Setting maximum ination pressure ........................................ 23
Taking fewer than three measurements .................................. 24
Skipping the countdown time ................................................... 24
Setting measurement interval times in «
» Mode ................... 25
Viewing the stored values .................................................... 26
Appendix
Rechargeable battery and power adapter ............................... 27
Troubleshooting ...................................................................... 28
Error messages.................................................................. 29-30
Safety, care, accuracy test and disposal ............................. 31-34
Technical specications .......................................................... 35
Table of Contents
6
Symbols
Documentation symbols
WARNING: The warning statements in this manual identify conditions or
practices that could lead to illness, injury
, or death.
CAUTION: The caution statements in this manual identify conditions or
practices that could result in damage to the equipment or other property,
or loss of data.
Consult operating instructions/directions for use (DFU). A copy of the
DFU is available on this web site. A printed copy of the DFU can be
ordered from Welch Allyn for delivery within 7 calendar days.
Helpful notes.
Shipping, storing, and environment symbols
Fragile; handle with care
Recovery/Recyclable
Recycle the product separate from other disposables
Keep dry
Control symbols
Power on/Power off
Start/Stop
Memory
Adjust measurement interval times
Adjust maximum ination pressure
1x
Single Reading Mode
Manual Mode
Three Reading Average Mode
EN
7
Welch Allyn ProBP 2400 digital blood pressure device
WARNING The information in this directions for use is a comprehensive guide to the
operation of ProBP 2400. For best results, read this directions for use thoroughly
before using the device.
WARNING The device is intended for use only in environments with clinician
supervision.
WARNING The device is designed for medical clinician use. Although this directions
for use may illustrate medical spot-check techniques, only a trained clinician
should use this device.
WARNING The device is not intended for use during patient transport.
WARNING Fire and explosion hazard. Do not operate the device in the presence
of a ammable anesthetic mixture with air, oxygen, or nitrous oxide; in oxygen
enriched environments.
WARNING Every three months, inspect the blood pressure cuff and other
accessories for fraying or other damage. Replace as necessary.
WARNING Inaccurate measurement risk. Do not use the device on patients who
are connected to heart/lung machines.
WARNING Electric shock hazard. Do not open the device or attempt repairs. There
are no user-serviceable parts inside ProBP 2400 other than battery replacement.
Only perform routine cleaning and maintenance procedures specically
described in this directions for use. Inspection and servicing of internal parts shall
only be performed by qualied service personnel.
WARNING The device complies with applicable domestic and international
standards for electromagnetic interference and should not present problems to
other equipment or be affected by other devices. As a precaution, avoid using the
device in close proximity to other equipment.
General warnings and cautions
Miscellaneous symbols
Meets essential requirements of European Medical Device Directive
93/42/EEC
Regulatory Affairs Representative
Type BF applied parts
Manufacturer
8
WARNING Welch Allyn is not responsible for the integrity of any mounting
installation. Welch Allyn recommends that customers contact their Biomedical
Engineering Department or maintenance service to ensure professional
installation for safety and reliability of any mounting accessory.
WARNING The device may not function properly if dropped or damaged. Do not use
the device if you notice any signs of damage. Qualied service personnel must
check any device that is dropped or damaged for proper operation before putting
the device back into use.
WARNING Defective batteries can damage the device. If the battery shows any
signs of damage, leakage, or cracking, it must be replaced immediately, and only
with a battery recommended for or supplied with the device.
WARNING Improper disposal of batteries may create an explosion or contamination
hazard. Never dispose of batteries in refuse containers. Do not dispose of the
battery in re. Always recycle batteries according to local regulations.
WARNING Improper handling of the battery can lead to heat generation, smoke,
bursting, or re.
WARNING Do not disassemble, modify, or solder the battery.
WARNING For proper patient electrical isolation and battery charging, use only the
provided external power supply to charge the device.
WARNING Electric shock hazard. Before cleaning the device, disconnect the power
cord from the power source and the device.
WARNING Take care to prevent water or other uid from entering any connectors
on the device. Should this occur, dry the connectors with warm air. Check the
accuracy of all operating functions.
Avoid simultaneously connecting patients to the device and high
frequency surgical equipment.
CAUTION The device is not heat-resistant. Do not autoclave.
CAUTION Use the device within stated operating temperature ranges. The device
will not meet performance specications if used outside these temperatures
ranges.
CAUTION Always unplug the external power source from the outlet before moving
the device to a new location.
CAUTION Use only Welch Allyn approved accessories. Use of unapproved
accessories with the device can affect patient and operator safety, and can reduce
product performance and accuracy.
CAUTION Federal law restricts this device to sale by or on the order of a physician or
licensed health care provider
EN
9
Welch Allyn ProBP 2400 digital blood pressure device
NIBP (non-invasive blood pressure) warnings and cautions
WARNING ProBP 2400 is not intended to measure blood pressure on children
younger than 3 years of age.
WARNING Do not compress the blood pressure hose or cuff. This may cause system
errors or patient safety risks to occur.
WARNING Inaccurate measurement risk. Do not use the device on patients who are
experiencing convulsions or tremors.
WARNING Use only blood pressure cuffs and hoses listed as approved accessories
to ensure safe and accurate of blood pressure measurements.
WARNING Patient injury risk. When several blood pressure measurements are
taken on the same patient, regularly check the cuff site and extremity for possible
ischemia, purpura, and/or neuropathy.
WARNING Do not allow a blood pressure cuff to remain on the patient more than
5 minutes when inated above 15 mmHg. Excessive cuff tightness may cause
venous congestion, peripheral nerve injury, discoloration of the limb, and patient
distress.
WARNING Patient injury risk. Never install Luer Lock connectors on Welch Allyn
blood pressure tubing. Using these connectors on blood pressure cuff tubing
creates the risk of mistakenly connecting this tubing to a patient’s intravenous line
and introducing air into the patient’s circulatory system.
WARNING NIBP measurements may be inaccurate in the presence of excessive
motion artefact. Minimize extremity and cuff motion during blood pressure readings.
WARNING The position and physiologic condition of the subject can affect a blood
pressure reading.
CAUTION If the blood pressure cuff is not at heart level, note the difference in reading
due to the hydrostatic effect.
CAUTION Proper blood pressure cuff size and placement is essential to the accuracy
of the blood pressure determination. See Blood pressure cuff selection for sizing
information.
CAUTION The blood pressure cuff must be properly positioned to ensure blood
pressure accuracy and patient safety. Wrapping the cuff too loosely (preventing
proper ination) may result in inaccurate blood pressure readings.
CAUTION The patients should be comfortably seated, legs uncrossed, feet at on
the oor, back and arm supported. The middle of the cuff should be at the level of
the heart.
CAUTION Recommend the patients to sit down and relax for at least 5 minutes prior
to the measurement and not talk during the measurement procedure.
10
3 sec.1x 3x3 sec.
On/Off Switch
Power Adapter Socket
Battery Compartment
Cuff Socket
Recharge Indicator
Display
M Button (Memory)
Mode Switch
Start/Stop Button
Product description
Name of parts
EN
11
Welch Allyn ProBP 2400 digital blood pressure device
EN
Display
1x
3x
260
240
220
200
180
160
140
AUTO
34.7
32.0
29.3
26.7
24.0
21.3
18.7
AUTO
3x
M1 M2 M3
PAM MAP
/ MIN
SYS
DIA
mmHg kPa
M
Systolic Value
Diastolic Value
Memory
Mean Arterial Pressure
Pulse Rate
Pulse Detected
Mode Selection:
Single Reading,
Manual, Three Reading Average
Power
Low Battery
Unit of Measure
Auto Inflation
Maximum Inflation Pressure
Rest/Count Down
Irregular Heart Beat (IHB)
Power
Low Battery
Mode Selection:
Single Reading,
Manual, Three Reading Average
Unit of Measure
Maximum Ination Pressure
Rest/Count Down
Auto Ination
Memory
Systolic Value
Diastolic Value
Mean Arterial Pressure
Irregular Hear Beat (IHB)
Pulse Rate
Pulse Detected
12
DESCRIPTIONPART NO.
TITLE:
AW, CD LABEL, DFU, PROBP 3400, ML 1.03
SHEET 1 OF 1
SCALE : 1:1
VER
CHANGE NUMBER/APVD
DESCRIPTION OF CHANGE
A RELEASE TO PRODUCTION D*
D* SEE SAP DIR FOR CHANGE NUMBER, APPROVER NAME AND DATE OF APPROVAL
408263
103701
NOTES:
1. USE CMYK COLOR MODE AND WHITE FLOOD FOR PRINTING ONTO THE CD/DVD SURFACE.
2. THE CD SHAPE LINES ARE FOR REFERENCE ONLY AND SHOULD NOT BE PRINTED.
3. THE TITLE TEXT IS C-58, M-17, Y-0, K-46, EQUAL TO PMS 5405C DARK BLUE.
4. INFORMATION TEXT COLOR IS BLACK IN CMYK VALUES C-50, M-50, Y-50, K-100.
5. WELCH ALLYN WORDMARK IS GREEN C-100, M-0, Y-91, K-6, AND BLUE C-100, M-72, Y-0, K-38.
6. UNIVERS FONTS IN BOLD AND REGULAR ARE USED. LANGUAGE FONT IS ARIAL UNICODE MS
REGULAR. FONT TYPE, ATTRIBUTES, AND COLOR SHALL BE CLEARLY LEGIBLE.
Microlife Corporation
9F, 431, RuiGuang Road, NeiHu,
Taipei, 11492, Taiwan, R.O.C.
Microlife AG
Espenstrasse 139
9443 Widnau, Switzerland
Distributed by Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls, NY 13153
Welch Allyn
ProBP
TM
2400
digital blood pressure device
Directions for use
Version 1.01
DESCRIPTIONPART NO.
TITLE:
AW, CD LABEL, DFU, PROBP 3400, ML 1.03
DRAWING NO:
SHEET 1 OF 1
DATE: 2014-12-24
DRW BY: Microlife
SCALE : 1:1
VER
CHANGE NUMBER/APVD
DESCRIPTION OF CHANGE
A RELEASE TO PRODUCTION D*
D* SEE SAP DIR FOR CHANGE NUMBER, APPROVER NAME AND DATE OF APPROVAL
408263
CD, DFU, PROBP 2400, ML 1.01
103701
DFU, PROBP 2400
NOTES:
1. USE CMYK COLOR MODE AND WHITE FLOOD FOR PRINTING ONTO THE CD/DVD SURFACE.
2. THE CD SHAPE LINES ARE FOR REFERENCE ONLY AND SHOULD NOT BE PRINTED.
3. THE TITLE TEXT IS C-58, M-17, Y-0, K-46, EQUAL TO PMS 5405C DARK BLUE.
4. INFORMATION TEXT COLOR IS BLACK IN CMYK VALUES C-50, M-50, Y-50, K-100.
5. WELCH ALLYN WORDMARK IS GREEN C-100, M-0, Y-91, K-6, AND BLUE C-100, M-72, Y-0, K-38.
6. UNIVERS FONTS IN BOLD AND REGULAR ARE USED. LANGUAGE FONT IS ARIAL UNICODE MS
REGULAR. FONT TYPE, ATTRIBUTES, AND COLOR SHALL BE CLEARLY LEGIBLE.
English, , Deutsch, Español, Português, Français, Italiano,
Nederlands, Dansk, Norsk, Svenska, Suomi, Русский, Polski,
Türkçe, 한국어
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
ProBP 2400 components and accessories
1 X ProBP 2400
1 X Adult (22cm~32cm)
1 X Large adult (32cm~42cm)
1 X CD Directions For Use
3 sec.1x 3x3 sec.
1 X Power Adapter
4 X Power Plugs (US, Europe, UK,
Australia)
(Input: 100-240V~50/60Hz 0.5A
Output: +7.5V 1.5A)
EN
13
Welch Allyn ProBP 2400 digital blood pressure device
Initial set up
Attaching the power plug to the power adapter
Select a proper power plug and attach to the power adapter as
shown below.
Charge the battery completely
When using for the rst time, charge the battery until the
recharge indicator on the device turns to green.
Refer to page 27 for the section of “Using a power adapter“.
Refer to page 27 for the section “Rechargeable Battery“
14
Selecting units of measure
1) Make sure the device is switched off.
2) Press and hold the M Button and then turn on the
power with the On/Off Switch.
3) Release the M Button when backlight illuminates.
4) Press the Start/Stop Button to select the preferred
pressure unit (mmHg or kPa).
5) Press M Button to conrm the selection.
Units
MAP or
PAM
1
mmHg MAP
2
mmHg PAM
3
kPa MAP
MAP is the abbreviation of Mean Arterial Pressure.
PAM or MAP is chosen depending on language preference.
Initial set up (cont.)
Factory default setting
EN
15
Welch Allyn ProBP 2400 digital blood pressure device
Each cuff is provided with 130 cm air tube.
Use only cuffs provided by Welch Allyn!
Contact Welch Allyn or its authorized distributor to purchase
cuffs.
Selecting the correct cuff
A variety of different cuff sizes are available. Adult and Large
Adult cuffs are provided with the device. Use the cuff marker
to select the cuff size that best matches the circumference of
the patient’s upper arm.
Adult and Large Adult cuffs are included as standard
accessories.
Reusable ProBP 2400
Blood Pressure Cuff
14-22cm
CHILD
www.welchallyn.com
!
Reusable ProBP 2400
Blood Pressure Cuff
32-52cm
LARGE ADULT LONG
www.welchallyn.com
!
LATEX
LATEX PVC DEHP
LATEX PVC DEHP
Reusable ProBP 2400
Blood Pressure Cuff
32-42cm
LARGE ADULT
www.welchallyn.com
!
LATEX
LATEX
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
www.welchallyn.com
!
LATEX
Child
Adult
Large Adult
Large Adult
Long
Before using the device
Cuff Size Circumference
(cm)
Circumference (inch)
Child 14 - 22 5.5 - 8.7
Adult 22 - 32 8.7 - 12.6
Large Adult 32 - 42 12.6 - 16.5
Large Adult
Long
32 - 52 12.6 - 20.5
16
Fitting the cuff properly
1) Place the cuff over the upper arm so that the air tube
and artery mark arrow point towards the lower arm.
The artery mark on the cuff must be placed over the
brachial artery.
2) Lay the cuff on the arm. Make sure that the lower edge
of the cuff lies approximately 2 to 3 cm ( ¾ to 1 inch)
above the elbow.
2-3 cm
(3/4 to 1 inch)
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
3) Wrap and tighten the cuff around the arm.
4) Leave free space with the size of 2 ngers between
the arm of the patient and the cuff. Excessive tightness
may cause venous congestion and discoloration of
the limb. If the cuff is wrapped too loosely, it cannot
be inated properly and the measured values may be
inaccurate. Remove all clothing covering or constricting
the measurement arm. Clothing may interfere with
measurement accuracy.
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
Before using the device (cont.)
EN
17
Welch Allyn ProBP 2400 digital blood pressure device
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
www.welchallyn.com
!
LATEX
5) Cuffs that do not t properly may lead to inaccurate
readings. Use a different size cuff if the range index at the
end of the cuff does not fall into the range specied by the
range stripes.
The range index of the cuff
should fall into this range.
18
Taking measurements in 1x, Manual and Mode
Set maximum ination pressure
Select desired maximum ination pressure or "AUTO".
Select an operation mode
There are three measurement modes that can be used.
Slide the switch to select 1x (standard single measurement),
Manual or Mode (automatic three reading average).
Turn on the power
Turn on the device by pressing the On/Off Switch at the back
of the device to the ON position.
Connect the cuff to the device
Connect the cuff to the device by inserting the cuff connector
into the cuff connector socket.
3 sec.1x 3x3 sec.
1x 3x
1x 3x
1x 3x
Refer to page 22 for the section of “Set maximum ination
pressure“
EN
19
Welch Allyn ProBP 2400 digital blood pressure device
«1x» Mode (standard single measurement)
Select «1x» Mode then press the «Start/Stop» Button
to perform a single blood pressure measurement. The
measurement reading is displayed and saved after the
measurement.
1x 3x
1x 3x
1x 3x
20
«Manual» Mode
Select «Manual» Mode if auscultatory blood pressure
measurement is preferred above using the oscillometric
method. In «Manual» Mode, the device serves only as a
pressure gauge. No oscillometric measurements will be
taken. The user can hear the systolic and diastolic Korotkoff
sounds by means of a stethoscope placed over the Brachial
artery.
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
Taking measurements in 1x, Manual and Mode (cont.)
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
Start ination – Press the Start/Stop Button to start ination
of the cuff.
When maximum ination pressure is reached, the ProBP
2400 will automatically begin linear deation at a rate of
3mmHg/sec.
Re-inate –Push and Hold the M Button during deation
to re-inate for as long as the button is held up to a max
of 299mmHg. Release the button to continue deation.
Exceeding 299mmHg will result in an immediate release of
cuff pressure and a 'HI' Error message.
EN
21
Welch Allyn ProBP 2400 digital blood pressure device
Stand by
The user can manually select up measurement interval times
of 15, 30, 45 or 60 seconds in
Mode. (Please refer to special
functions section page 21 “Setting measurement interval times”).
The 60 second wait period before the rst measurement is not
adjustable but may be bypassed by pressing the Start/Stop
Button a second time. This will start the rst reading immediately.
« » Mode (automatic three reading average)
Select « » Mode then press the «Start/Stop» Button to
perform automatic triple blood pressure measurements to
determine a three reading average.
The countdown time before the rst measurement is set at 60
seconds.
The interval times between measurements is user adjustable to
15, 30, 45 or 60 seconds. The average measurement reading is
displayed and saved after the measurements are complete.
15, 30, 45,
or 60 sec.
15, 30, 45,
or 60 sec.
1 2
3
60 sec.
1x 3x
When the cuff pressure reaches 20mmHg during the
deation cycle, the remaining pressure is vented and the
ProBP 2400 goes into the Stand by mode.
Push the «Start/Stop» at any time to start fast deation and
set the ProBP 2400 to Stand by.
Take note – Take note of the systolic and diastolic values in
the same manner as performed with sphygmomanometer
measurements.
Set to ‘Stand by’ – The device can be set to ‘‘Stand by” by
pressing the «Start/Stop» Button without turning off the
power. The device will automatically switch to ‘Stand by’ if
there is no operation for one minute.
22
Special Functions
MAP (Mean Arterial Pressure)
The device measures the true mean arterial pressure
(MAP) of the patient. Each measurement includes a single
MAP value. The MAP value will always be displayed
together with the systolic and diastolic blood pressure
value.
Irregular heartbeat detector in «1x» Mode
The device detects irregular heartbeat in «1x» Mode.
The irregular heartbeat symbol shows up if during a
blood pressure measurement the heart rate has varied
by more than 25%. In case of an irregular heartbeat the
blood pressure measurement might be affected. It is
recommended to repeat the measurement or to choose
the «Manual» Mode for verication.
The Mean Arterial Pressure, MAP is determined from
the maximum peak of the oscillometric envelope curve.
EN
23
Welch Allyn ProBP 2400 digital blood pressure device
The ination pressure selection (mmHg/kPa) can be done in 1x,
, or Manual Mode. The default setting is “AUTO”
When set to AUTO, the ProBP 2400 will automatically determine
the maximum ination pressure as it inates the cuff.
Set maximum ination pressure
1) Press and hold the M Button for 3 seconds until the column
with pressure values ashes.
2)
Press the Start/Stop Button to select the preferred pressure
value (after value ‘260’ has been reached the next selection
option will be “AUTO” at the bottom of the list)
3) Press the M Button to conrm the selected value.
Push and hold for 3
seconds...
3 sec.
3 sec.
If the maximum ination pressure selected (or the
maximum ination pressure as determined while in
AUTO mode) is not adequate to determine systolic
pressure, the device will reinate to a pressure 30
mmHg higher than the previous ination pressure and
attempt another cycle. This can be repeated increasing
the maximum ination by 30 mmHg each time. If a
maximum ination pressure of 280 mmHg is reached,
and the ProBP 2400 is unable to determine a blood
pressure, an error code will be displayed.
It is recommended at this point to switch to the manual
mode and determine blood pressure with a stethoscope
using traditional Korotkoff method. (See «Manual»
Mode)
24
Taking fewer than three measurements in
«
» Mode
The measurement sequence can be stopped at any time
during the measurement sequence by pressing the Start/
Stop Button. The device enters ‘Stand by’ and remaining
measurements are cancelled. Data from the measured
blood pressure can be viewed by pushing the M Button.
Skipping countdown time in « » Mode
The countdown before and between measurements in « »
Mode can be skipped by pressing the Start/Stop Button. When
the Start/Stop Button is pressed during countdown, the device
will immediately begin the next measurement.
The device can be set in ‘Stand by’ Mode by pressing the Start/
Stop Button after completion of measurements. The device will
automatically switch to ‘Stand by’ Mode if left unattended for 1
minute.
Cancel remaining
measurements at
any time during
the measurement
sequence.
Skip the countdown time and
begin the measurement.
Stand by
Special Function (cont.)
EN
25
Welch Allyn ProBP 2400 digital blood pressure device
Setting measurement interval times in
«
» Mode
The default measurement interval time is 60 seconds. The
interval times can be set as 15, 30, 45 or 60 seconds.
1) Press and hold the Start/Stop Button for 3 seconds.
Press and hold for 3
seconds...
2) Press the M Button to adjust the measurement interval
time, then press the Start/Stop Button to conrm. The
device will go back to ‘Stand by’.
15 30
60
45
3 sec.
3 sec.
26
Viewing stored values
1) In «1x» Mode
2) In «
» Mode
Press the M Button to reveal the average of the triple
measurements. Continue pressing the M Button to
review individual measurements.
Viewing stored values
The device only stores blood pressure values of the last
measurement procedure in «1x» and « » Mode. Press the
M Button to review the stored readings when the device is
in 'Stand by' mode.
The device stores only the last measurement completed in
«1x » Mode and the last three measurements completed in
«
» Mode.
EN
27
Welch Allyn ProBP 2400 digital blood pressure device
Rechargeable Battery
The ProBP 2400 features a built-in, rechargeable Ni-MH
battery pack that provides up to 600~700 measurement
cycles on a full charge. The battery can be recharged with
the power adapter provided. The empty battery indicator is
displayed when the battery is low.
Rechargeable battery and power adapter
Using a power adapter
Only use the adapter supplied with ProBP 2400 to recharge
the device.
1
) Plug the adapter cable into the power sock et of the device.
2) Plug the adapter plug into the wall socket. The battery
will be recharged as long as the device is attached to an
AC power source. After the battery is fully recharged, the
charging will stop. No battery power will be used as long
as the adapter is plugged in. The battery has to remain in
the ProBP 2400 also when using the AC power.
3) If the battery starts losing capacity, contact your local
dealer for replacement battery. The battery is user
replaceable.
3 sec.1x 3x3 sec.
When using for the rst time,
charge the battery until the
recharge indicator turns to
green.
The orange recharge
indicator means that the
recharge is in progress.
A green recharge indicator
means that the recharge is
completed.
28
Troubleshooting
Problem Possible cause Solutions
No power (No LCD display) Power supply is not properly
plugged in
Plug the power supply into the wall socket.
Battery is fully discharged Recharge the rechargeable battery by plugging in the
power supply.
Cuff does not inate properly Loose connection of the tube Make sure the tube of the cuff is securely connected to the
device.
Leakage of the tube / bladder Check for cracks on the tube or the bladder. Replace the
blood pressure cuff if necessary.
No result displayed after
measurements
Device is in Manual Mode
Switch to «1x» or «
» Mode and repeat the
measurements.
EN
29
Welch Allyn ProBP 2400 digital blood pressure device
If an error occurs during a measurement, the
measurement is interrupted and an error message
«Err» is displayed.
Error Description Potential cause and
remedy
«Err 1» Signal too
weak
The pulse signals on the
cuff are too weak. Re-
position the cuff and repeat
the measurement.
«Err 2» Error signal During the measurement,
error signals were detected
by the cuff caused, for
instance, by movement or
muscle tension. Repeat
the measurement keeping
patient’s arm still.
Error messages
30
«HI» Pulse rate or
cuff pressure
too high
The pressure in the cuff
is too high (over 300
mmHg) OR the pulse is too
high (over 200 beats per
minute). Have the patient
relax for 5 minutes and
repeat the measurement.
«LO» Pulse too low The pulse is too low (less
than 40 beats per minute).
Repeat the measurement.
«Err 3» No pressure in
the cuff
An adequate pressure
cannot be generated in
the cuff. A leak may have
occurred. Replace the
blood pressure cuff if
necessary. Repeat the
measurement.
«Err 5» No valid
results
The measuring signals are
inaccurate therefore no
result can be displayed.
Switch to Manual
Mode and determine
blood pressure with
a stethoscope using
traditional Korotkoff
method.
EN
31
Welch Allyn ProBP 2400 digital blood pressure device
Protect the device from water and moisture
Protect the device from extreme heat and cold
Protect device from impact and drops
Protect the device from direct sunlight
Avoid proximity to electromagnetic elds, such as
those produced by mobile phones
Never open device
Safety, care, accuracy test and disposal
Safety and protection
This device may be used only for the purpose described
in this directions for use manual. The device comprises of
sensitive components and must be treated with care. The
manufacturer cannot be held liable for damage caused by
incorrect application.
Observe the storage and operating conditions described in
the “Technical specications” section of this directions for
use manual.
Only activate the pump when the cuff is connected to
the device.
Do not use the device if you think it is damaged or if
anything appears unusual.
Read further safety instructions in the individual
sections of the instruction manual.
32
Do not iron the cuff!
Cleaning the cuff
Remove the bladder. Fold and place the cuff cover inside a
washing bag. Wash the cuff cover with warm water (43°C;
110°F) and a mild detergent in the washing machine.
Pasteurization: wash the cuff cover in 75°C (167°F) hot
water for 30 minutes.
Air dry the cuff. DO NOT iron the cuff cover
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
www.welchallyn.com
!
LATEX
Device care
Use a soft cloth with one of the following recommended
cleaning solutions to wipe the exterior of the device:
1) Mild soap and water
2) Hydrogen peroxide solution (3% diluted with water)
3) Sodium hypochlorite solution (1:10 dilution of
household chloride bleach in water)
EN
33
Welch Allyn ProBP 2400 digital blood pressure device
Transducer accuracy test
(Calibration Verication)
We recommend the device to be tested for accuracy every 2
years or after mechanical impact (e.g. been dropped).
1) Setup for accuracy test
2) Push and hold “Start/Stop” Button and turn the power on
then release the “Start/Stop” Button. Wait until ““CA 0”,
then “0 0“ is displayed.
350~500 ml
Ves sel
35 0~500 ml
Ves sel
Air release
valve
C alibrated
referen ce
manom e ter
ProBP 2400
T-pieces
Squeeze bulb
(or Pressure source)
ProBP 2400
34
Disposal
Batteries and electronic instruments must
be disposed of in accordance with the locally
applicable regulations, and not as domestic
waste.
3) Pump the pressure to nearly 100mmHg. Compare the
pressure displayed on the screens of the device and
the reference manometer. For example, the “100 .9”
displayed on the device stands for “100.9 mmHg”.
4) Pump the pressure to nearly 200mmHg. Compare the
pressure displayed on the screens of the device and
the reference manometer. For example, the “201.0”
displayed on the device stands for “201.0 mmHg”.
5) Pump the pressure to nearly 300mmHg. Compare the
pressure displayed on the screens of the device and
the reference manometer. For example, the “301.2”
displayed on the device stands for “301.2 mmHg”.
6) If the difference between the device and the reference
manometer at any calibration point exceeds ±3mmHg
plus the stated accuracy of the reference manometer,
you may contact Welch Allyn to obtain calibration
service.
EN
35
Welch Allyn ProBP 2400 digital blood pressure device
Technical specications
Operation temperature/
humidity:
10 to 40 °C (50 to 104 °F)
15 - 90 % relative maximum humidity
Storage temperature/
humidity:
-20 to 55 °C (-4 to 131 °F)
15 - 90 % relative maximum humidity
Weight: 800 g (including rechargeable battery pack)
Dimensions: 200 x 125 x 90 mm
Measuring method: Oscillometric,
Systolic blood pressure = K1
Diastolic blood pressure = K5
Measurement range: 60 - 255mmHg - systolic blood pressure
30 - 200mmHg - diastolic blood pressure
40 - 200 beats per minute - pulse
Cuff pressure display: Range: 0 - 299 mmHg
Resolution: 1 mmHg
Static accuracy: pressure within ± 3 mmHg
Pulse accuracy: ±5 % of the readout value
Power source: Rechargeable battery pack;
4.8V 2400 mAh (optional 4.8V 3500mAh)
Mains power supply DC 7.5V, 1.5 A
Reference to Standards: Device corresponds to the requirements of the
standard for non-invasive blood pressure monitor.
IEC 60601-1: 2005+A1:2012
IEC 60601-1-2:2014
AN O 81060-2
SI/AAMI/IS
AAMI / ANSI / IEC 80601-2-30
Electromagnetic
Compatibility:
Device fullls the stipulations of the standard IEC
60601-1-2.
The stipulations of the EU Directive 93/42/EEC for Medical
Devices Class IIa have been fullled.
Type BF applied part
Welch Allyn reserves the right to alter technical specications without prior written notice.
ProBP 2400
ProBP 2400 Digital Blood Pressure Monitor
www.welchallyn.com
BOM number:
Color: CMYK
Material: 80P Coated Paper
9010962400
DIGITAL BLOOD PRESSURE DEVICE, ML
Distributed by Welch Allyn Inc.
4341 State Street Road
Skaeateles Falls, NY 13153
Microlife Corporation
9F, 431, RuiGuang Road, NeiHu,
Taipei, 11492, Taiwan, R.O.C.
Microlife AG
Espenstrasse 139
9443 Widnau, Switzerland
PROBP 2400 DIGITAL DEVICE
#
R
x ONLY
LATEX
GTIN
(01)00732094211078
(BOM number)
(Material: 80P Coated Paper)
(Model: ProBP 2400)
Prepared by:
Approved by:
Tolerances: +0.5, -0.5
(Uint: mm)
(21)41140001
7d
welchallyn.com
ProBP 2400
ProBP 2400 Digital Blood Pressure Monitor
www.welchallyn.com
BOM number:
Color: CMYK
Material: 80P Coated Paper
9010962400
DIGITAL BLOOD PRESSURE DEVICE, ML
Distributed by Welch Allyn Inc.
4341 State Street Road
Skaeateles Falls, NY 13153
Microlife Corporation
9F, 431, RuiGuang Road, NeiHu,
Taipei, 11492, Taiwan, R.O.C.
Microlife AG
Espenstrasse 139
9443 Widnau, Switzerland
PROBP 2400 DIGITAL DEVICE
#
R
x ONLY
LATEX
GTIN
(01)00732094211078
(BOM number)
(Material: 80P Coated Paper)
(Model: ProBP 2400)
Prepared by:
Approved by:
Tolerances: +0.5, -0.5
(Uint: mm)
(21)41140001
7d
welchallyn.com
ETL CLASSIFIED
5000497
CONFORMS TO AAMI STD ES 60601-
1, IEC STD. 60601-1-6, IEC STD.
80601-2-30
CERTIFIED TO CSA STD C22.2 NO.
60601-1, NO. 60601-1-6, NO. 80601-2-30
3
6
Welch Allyn will warranty the ProBP 2400 device to be free of defects in material and workmanship and to perform in
accordance with manufacturer specications for a period of two years from the date of purchase from Welch Allyn or its
authorized distributors or agents. Accessories are covered for a period of one year from date of purchase from Welch
Allyn or its authorized distributors or agents.
The warranty period shall start on the date of purchase. The date of purchase is: 1) the invoiced ship date if the device
was purchased directly from Welch Allyn, 2) the date specied during product registration, 3) the date of purchase of
the product from a Welch Allyn authorized distributor as documented from a receipt from said distributor.
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or maintenance contrary to
labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents.
The product warranty is also subject to the following terms and limitations:
Shipping cost to return a device to a Welch Allyn Service center is not included.
A service notication number must be obtained from Welch Allyn prior to returning any products or accessories to
Welch Allyn’s designated service centers for repair. To obtain a service notication number, contact Welch Allyn
Technical Support.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT
NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. WELCH ALLYN’S OBLIGATION UNDER THIS WARRANTY IS LIMITED TO REPAIR OR
REPLACEMENT OF PRODUCTS CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY
INDIRECT OR CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY THE
WARRANTY.
WARRANTY
EN
3
7
Welch Allyn ProBP 2400 digital blood pressure device
IB ProBP 2400 EN_FDA 1717
© 2014 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser
of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other
use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.
For the addresses and phone numbers of all Welch Allyn regional locations, visit: welchallyn.com/locations.