A high purity toluene sample was also run to demonstrate the large, automated
response range of the Intuvo 9000 FID (Figure 2). Note that the detector is able
to respond to peaks from 0.16 pA to 13,500 pA. This represents a concentration
difference from 18ppm to 99.97 weight %.
51015202530
8
10
12
14
16
18
20
min10.8 10.9 11.0 11.1 11.2 11.3
pA
0
2,000
4,000
6,000
8,000
10,000
12,000
14,000
Toluene
Ethylbenzene
Non-aromatics
0.16 pA
13,500 pA
C9 + aromatics
Figure 2. D7504 Analysis of purified toluene. Very large components (99.97 %) and very small components
(18 ppm) were measured in a single run.
Table 1. Results for 10 D7504 analyses of mixed xylenes.
Run
Nonaromatics
wt %
Toluene
wt %
Ethylbenzene
wt %
p-Xylene
wt %
m-Xylene
wt %
Cumene
wt %
o-Xylene
wt %
C9+ Aromatics
wt %
1 0.1982 0.0101 16.84 21.05 46.43 0.0221 15.29 0.1667
2 0.1991 0.0100 16.83 21.05 46.43 0.0219 15.30 0.1662
3 0.1986 0.0100 16.84 21.05 46.43 0.0218 15.29 0.1646
4 0.1994 0.0100 16.83 21.05 46.43 0.0219 15.30 0.1595
5 0.1984 0.0100 16.83 21.04 46.43 0.0219 15.30 0.1623
6 0.1993 0.0101 16.84 21.04 46.43 0.0219 15.31 0.1651
7 0.2008 0.0100 16.83 21.04 46.43 0.0219 15.30 0.1632
8 0.1998 0.0101 16.84 21.04 46.43 0.0219 15.30 0.1566
9 0.2005 0.0100 16.83 21.04 46.43 0.0219 15.30 0.1624
10 0.2005 0.0100 16.83 21.04 46.43 0.0219 15.31 0.1633
Mean 0.1995 0.0100 16.83 21.04 46.43 0.0219 15.30 0.1630
StdDev 0.00092 0.00005 0.004 0.004 0.003 0.00007 0.006 0.00307
ASTM SD 0.00700 0.01400 0.007 0.029 0.021 0.00003 0.010 0.00100